Please use this identifier to cite or link to this item:
http://repository.ipb.ac.id/handle/123456789/158506
Title: | Simulasi Monte Carlo Brakiterapi pada Kanker Superfisial dengan Sumber Co-60 |
Other Titles: | Monte Carlo Simulation Superficial Cancer Brachytherapy using Source Co-60 |
Authors: | Puspita, R. Tony Ibnu Sumaryada Wijaya Yani, Sitti Uzlifah, Maulida Janatu |
Issue Date: | 2024 |
Publisher: | IPB University |
Abstract: | Penelitian ini menggunakan simulasi Monte Carlo untuk membandingkan
kurva isodosis dan dose volume histogram pada fantom inhomogen yang diradiasi
dengan Cobalt-60. Sumber radioaktif ditempelkan dalam polyethylene. Ukuran
fantom, yaitu 10 × 10 × 10 cm3
. Penelitian ini menggunakan target berukuran 2 cm
× 2 cm × 3 mm dan 2 cm × 2 cm × 5 mm. Jarak target simulasi dengan sumber
radioaktif, yaitu 0 dan 3 mm. Hasil penelitian menunjukkan penurunan dosis pada
target seiring bertambahnya kedalaman (5 mm, 7 mm, 10 mm) karena atenuasi
radiasi. Faktor lain yang memengaruhi adalah densitas target, dimana target dengan
densitas lebih tinggi menyerap lebih banyak radiasi. Brakiterapi terbukti efektif
untuk target dangkal (5-7 mm) dengan distribusi dosis mencapai 20%-40% dari
dosis maksimum. Namun, efektivitasnya menurun pada target yang lebih dalam
(>10 mm) karena dosis yang diterima target berkurang hingga di bawah 10% dari
dosis maksimum. This study used Monte Carlo simulation to compare isodose curves and dose volume histograms on inhomogeneous phantoms irradiated with Cobalt-60. The radioactive source was attached to polyethylene. The size of the phantom was 10 × 10 × 10 cm3 . This study used 2 cm × 2 cm × 3 mm and 2 cm × 2 cm × 5 mm targets. The distance between the simulated target and the radioactive source was 0 and 3 mm. The results show a decrease in the dose to the target as the depth increases (5 mm, 7 mm, 10 mm) due to radiation attenuation. Another influencing factor is the target density, where higher-density targets absorb more radiation. Brachytherapy is effective for shallow targets (5-7 mm) with dose distribution reaching 20%-40% of the maximum dose. However, its effectiveness decreases in deeper targets (>10 mm) |
URI: | http://repository.ipb.ac.id/handle/123456789/158506 |
Appears in Collections: | UT - Physics |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
cover_G7401201011_ea4a090462f04d5cae96d245e750290c.pdf | Cover | 423.84 kB | Adobe PDF | View/Open |
fulltext_G7401201011_ebb173e86a9e4be588c10eb8da47df44.pdf Restricted Access | Fulltext | 1.39 MB | Adobe PDF | View/Open |
lampiran_G7401201011_b8a1e139bee844e6ad22f9add8b69e9f.pdf Restricted Access | Lampiran | 79.51 kB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.