Please use this identifier to cite or link to this item: http://repository.ipb.ac.id/handle/123456789/70368
Title: Linearitas dan Limit Deteksi Pasta Karbon Biosensor Arsen dengan Elektrode Pasta Karbon Termodifikasi Zeolit-Fe
Authors: Pradono, Dyah Iswantini
Purwaningsih, Henny
Ghozali, Ali Aulia
Issue Date: 2014
Abstract: Detection of arsenic using biosensor based on enzyme inhibition could be a routine alternative analysis method of aquatic environmental quality. Biosensors were fabricated from carbon paste electrode modified by zeolite-Fe with pyruvate dehydrogenase (PDH) enzyme as bioreceptor which was inhibited its activity by arsenic. Measurement results were compared to atomic absorption spectroscopy (AAS) measurement. Linearity and LOD value of both methods were compared. The results showed optimum performance of biosensor was at pH 7.00, 33 °C, and at PDH concentration of 0.0142 U/mL. The best linearity was around 97.70% in range of 2.50-20 ppb, meanwhile the best LOD achieved as low as 3.78 ppb. Biosensor’s LOD was lower compared to AAS value of LOD measured as low as 6.52 ppb. Paired data test showed that there was no significant difference between the two methods at 95% confidence level. Reproducibility of fabricated biosensor was still low, hence it needs to enhance the biosensor performance
URI: http://repository.ipb.ac.id/handle/123456789/70368
Appears in Collections:UT - Chemistry

Files in This Item:
File Description SizeFormat 
G14aag.pdf
  Restricted Access
full text1.84 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.