Please use this identifier to cite or link to this item: http://repository.ipb.ac.id/handle/123456789/169482
Title: Validasi Metode Penetapan Kadar Tembaga dalam Konsentrat Tembaga menggunakan ICP-OES di Balai Laboratorium Bea dan Cukai
Other Titles: Method Validation for Copper Content Determination in Copper Concentrate using ICP-OES at Balai Laboratorium Bea dan Cukai
Authors: Soebrata, Betty Marita
Rinjani, Tivani
Issue Date: 2025
Publisher: IPB University
Abstract: Kadar tembaga dalam konsentrat tembaga perlu dilakukan untuk menjaga keadilan dan transparansi dalam sektor pertambangan. Kadar minimum tembaga dalam konsentrat berdasarkan Peraturan Menteri Perdagangan Republik Indonesia Nomor 11 Tahun 2024 tentang Kebijakan dan Pengaturan Ekspor yaitu sebesar =15%. Metode divalidasi agar memberikan hasil yang akurat. Validasi metode dilakukan dengan beberapa parameter validasi seperti linearitas, BD, BK, akurasi, presisi, ketahanan, dan ketangguhan. Hasil pengujian linearitas diperoleh persamaan garis y = 10952x+2979,8 dengan nilai koefisien korelasi (r) = 0,9999. BD sebesar 0,6330 ppm dan BK sebesar 1,9183 ppm. Uji akurasi yang diperoleh yaitu 100,10%. Uji presisi yang diperoleh yaitu nilai %SBR 0,96. Uji ketahanan dan ketangguhan dengan uji F dan uji t tidak memberikan hasil yang berbeda nyata. Metode ini valid sehingga dapat digunakan sebagai analisis rutin di BLBC. Kadar konsentrat yang didapatkan sebesar 32,51% yang artinya memenuhi syarat Peraturan Menteri Perdagangan Republik Indonesia Nomor 11 Tahun 2024.
The determination of copper content in copper concentrate was essential to maintaining fairness and transparency within the mining sector. According to the Regulation of the Minister of Trade of the Republic of Indonesia Number 11 of 2024 concerning Export Policies and Regulations, the minimum copper content in concentrate was set at =15%. To ensure accurate results, the analytical method was validated, encompassing parameters such as linearity, Limit of Detection (LOD), Limit of Quantitation (LOQ), accuracy, precision, robustness, and ruggedness. Linearity testing yielded a linear equation of y=10952x+2979.8 with a correlation coefficient (r) of 0.9999, indicating a strong linear relationship. The LOD was 0.6330 ppm and the LOQ was 1.9183 ppm. The accuracy test resulted in a recovery of 100.10%, while precision testing yielded a %RSD of 0.96%. Robustness and ruggedness tests showed no significant differences in results, confirming the method's reliability. Consequently, the method was considered valid for routine analysis at BLBC. The copper concentrate content obtained was 32.51%, which met the requirements stipulated in the Regulation of the Minister of Trade of the Republic of Indonesia Number 11 of 2024.
URI: http://repository.ipb.ac.id/handle/123456789/169482
Appears in Collections:UT - Chemistry Analysis

Files in This Item:
File Description SizeFormat 
cover_J0312211146_651ca88934584d76ae9dcad8808eac4d.pdfCover2.25 MBAdobe PDFView/Open
fulltext_J0312211146_6a61153bac0d4be1b9da4b2db1d55175.pdf
  Restricted Access
Fulltext2.82 MBAdobe PDFView/Open
lampiran_J0312211146_84da4e2b18d64259ab2ae1104481ed30.pdf
  Restricted Access
Lampiran2.23 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.