Pembuatan dan pencirian elektrode selektif ion Lu3+ (Lutetium) dengan Ionofor 4- odekanadioilbis-(1-fenil-3-metil-5-pirazolon)

Date
2011Author
Asri, Hasha Bwana
Saprudin, Deden
Pradono, Dyah Iswantini
Metadata
Show full item recordAbstract
The ion selective electrode (ISE) performance is affected by the type of ionophore used. This research developed an Lu3+ ISE with dodecanedioylbis-(1- phenyl-3-methyl-5-pirazolone) (H2DdBP) as ionophore which has been proved selective in Lu3+ liquid-liquid extraction. The best composition of ISE obtained from this research was 33% PVC, 57% acetophenone, 5% H2DdBP, and 5% oleic acid. Linear measurement was obtained with Nernst factor of 16.57 mV/decade at 10-3−10-7 M concentration range in very acidic situation and response time was 15 to 250 seconds. ISE still could be used at pH 4 and 7, with 5% Nernst factor decrease after 30 days usage. Other ions such as Cu2+, Cd2+, Pb2+, Ni2+, Ca2+, and La3+ did not affect the measurement of ISE.
Collections
- UT - Chemistry [1694]