Show full item recordAbstract
Barium strontium titanate (BST) is a photodiode as a ferroelectric material that has been developed in many application such as sensor technology. The purpose of this study is to design a simple measurement tool of glucose with BST thin film as the sensor component and the signal conditioning. The measurement tool test was done by measuring blood glucose levels ranging from 20 to 180 mg/dL, which is the range of glucose levels for person who has lack glucose to the excess glucose. The test results showed that the measurement tool could detect each addition of glucose levels by 1 mg/dL which would reduce the value of the output voltage of 0.0022 V. In addition, the test results showed that the response of the circuit approach to linear, the correlation coefficient (R2) is 0.9155 and 0.8931. The test results also showed good hysteresis as showed by the relatively same graphic plots.